Ambulance defibrillator – which functions are most important for the rescue team

avatar
21 October 2026
12 mins reading.
Contents:
Ambulance defibrillator

An ambulance defibrillator combines five groups of functions, from manual and semi-automatic modes, through monitoring with capnography and cardioversion with external pacing, to ECG teletransmission and resistance to operating conditions in the vehicle. Each of these stems from the qualifications of the paramedics, the principles of conducting resuscitation, or the requirements placed on ambulance equipment.

How an ambulance defibrillator differs from an AED

An ambulance defibrillator works in a manual mode, where the rhythm is assessed by the paramedic, and in a semi-automatic mode, where the analysis is taken over by the device, and additionally, it monitors the patient. A classic AED analyzes the rhythm independently and guides the user with voice prompts.

  1. Manual mode and conditions for its use. In manual mode, the paramedic reads the ECG recording, selects the energy, and decides on the shock. The manual mode should be used by paramedics who can recognize a cardiac arrest rhythm within about 5 seconds and deliver a shock with a pause in chest compressions of less than 5 seconds. This requires the device to have a clear ECG display on the screen, as well as fast charging and shock delivery.
  2. Semi-automatic mode in the team’s work. Advanced life support personnel must be proficient in using both an AED and a manual defibrillator. A device with both modes eliminates the need to carry two pieces of equipment.
  3. Transitioning from an AED to a manual defibrillator. Sometimes the team finds an AED already connected to the patient. They then follow the device’s prompts and, if possible, switch to the manual defibrillator during a two-minute cycle of compressions. When choosing equipment, it is useful to check how many steps it takes to switch modes.

Which defibrillator functions does a paramedic use independently

A paramedic can independently perform manual defibrillation based on an ECG or cardiac monitor reading, automated defibrillation, transcutaneous cardiac pacing in bradyarrhythmias, and cardioversion in tachyarrhythmias, with pacing and cardioversion applying to hemodynamically unstable patients. Their duties also include ECG recording and assessment, respiratory monitoring, and non-invasive hemodynamic monitoring. These qualifications arise from Annex No. 1 to the Regulation of the Minister of Health on emergency medical procedures.

Procedure from Annex No. 1Device function
Manual defibrillation based on an ECG or cardiac monitorManual mode, clear ECG display, energy selection
Automated defibrillationSemi-automatic mode with rhythm analysis and prompts
Transcutaneous cardiac pacing in bradyarrhythmiasExternal pacing module
Cardioversion in tachyarrhythmiasSynchronized shock with the R wave
ECG recording and assessmentECG recording and display
Respiratory monitoring and non-invasive hemodynamic monitoringSpO2, EtCO2, and blood pressure measurement

Manual and automated defibrillation

Both types of defibrillation fall within the paramedic’s qualifications. A device without either mode restricts the team in performing a procedure they are authorized to do, and the difference comes down to who assesses the rhythm. With manual, the paramedic does it; with automated, the device does it.

Cardioversion and external pacing

Electrical cardioversion is preferred in an unstable patient with a tachyarrhythmia, and the shock must be synchronized with the R wave on the ECG. Instability is evidenced by life-threatening signs such as shock with hypotension (e.g., systolic pressure below 90 mmHg), syncope, heart failure, or myocardial ischemia. Transcutaneous pacing acts as a bridge to transvenous pacing or replaces it when the latter is unavailable.

The paramedic’s qualification does not mean that every defibrillator in every ambulance has both modules. The range of functions for a specific model is described in the manufacturer’s documentation, and the minimum equipment list for an ambulance is based on the ordinance of the President of the National Health Fund and the contracting authority’s specifications.

Ambulance defibrillator functions that shorten pauses in chest compressions

All functions in this group serve one purpose, which is to maintain the continuity of compressions.

The pause around the shock should be less than 5 seconds, and compressions resume immediately after the shock. In practice, this consists of six elements:

  • The defibrillator charges during compressions, the shock occurs after a short pause, and compressions are resumed immediately after it. Charging before rhythm assessment is also acceptable if it does not prolong the pause. Defibrillating during manual compressions is not allowed, even with gloves on, as it poses a danger to the rescuer.
  • It is useful to have an ECG preview with compression artifacts filtered out, provided the device offers such filtering. When asystole is visible on the screen, there is no need to stop compressions to check the rhythm.
  • The first shock for a biphasic waveform should be at least 150 J, and for a biphasic pulsed waveform 130–150 J. After a failed shock, it is worth increasing the energy of subsequent ones, if the device allows it, and this capability should be verified with the manufacturer.
  • The initial pad placement is antero-lateral, with the lateral pad placed directly under the armpit in the mid-axillary line. After three failed shocks in refractory rhythms, a new set of pads is prepared in an antero-posterior position, and a spare pair should be kept close at hand. In a patient with an implantable cardioverter-defibrillator (ICD), the pad must be placed more than 8 cm away from the device.
  • An ambulance device should be suitable for both adults and children.
  • Mechanical compressions are considered when high-quality manual compressions are impractical or compromise the rescuer’s safety. A shock from a manual defibrillator can then be delivered without interrupting the mechanical compressions.

Monitoring in an ambulance defibrillator

An ambulance defibrillator is simultaneously a cardiac monitor and, apart from the shock, provides the team with data on heart rate, blood oxygenation, blood pressure, and carbon dioxide in the exhaled air.

ParameterPurpose in the team’s work
ECGRhythm assessment, the basis for the decision on a shock, cardioversion, and pacing; the number of leads (3 or 12) must be checked for a given model
SpO2Assessment of the patient’s oxygenation during the intervention and transport
Non-invasive blood pressure (NIBP)Identification of hemodynamic instability, including hypotension
EtCO2 (waveform capnography)Confirmation of endotracheal tube placement and assessment of compression quality
TemperatureMeasurement of the patient’s temperature; core temperature measurement is among the paramedic’s procedures
MetronomeMaintaining the rhythm of compressions

For an unstable patient, it is necessary to check how the device allows for setting the interval between non-invasive blood pressure measurements. The screen and alarms must be readable from the distance at which the entire team works, even in the cramped medical space of an ambulance.

Waveform capnography confirms the placement of the endotracheal tube, as a sustained EtCO2 tracing rules out esophageal intubation, and allows for the assessment of compression quality. Endotracheal intubation should only be performed by rescuers with a high success rate and continuous waveform capnography. A paramedic performs intubation in sudden cardiac arrest independently, so an EtCO2 module is a practical requirement here. An increase in EtCO2 during resuscitation may indicate a return of spontaneous circulation, but it is not a reason to stop compressions on its own. The return of circulation is evidenced only by a combination of signs, for example, consciousness, purposeful movements, and a sudden rise in EtCO2, while a low EtCO2 value alone does not justify terminating resuscitation. The DEFIGARD Touch 7 by Schiller features diagnostic ECG with interpretation, blood pressure measurement, Masimo Rainbow SpO2, EtCO2 measurement, temperature, and a metronome.

ECG teletransmission and data recording from the ambulance defibrillator

ECG teletransmission allows a doctor to assess the recording remotely and has legal implications. A paramedic can administer clopidogrel, prasugrel, or ticagrelor only after ECG teletransmission and consultation with the doctor evaluating the reading.

If the defibrillator records the reading, whether the team can use this authorization depends on its transmission module. The DEFIGARD Touch 7 has direct data transmission, and the details must be arranged with the manufacturer.

The doctor views the transmitted reading, consults the decision with the paramedic, and only then does the paramedic administer the drug. The reading must reach the doctor in an assessable form; merely recording it on the device is not enough.

Before purchasing, four things must be determined:

  • what format the device uses to save and transmit the ECG,
  • how many leads the recording includes (3 or 12),
  • what system on the receiver’s end the transmission goes to,
  • how to read the data after the intervention is over.

Data recording also serves a purpose after the intervention. A resuscitation debriefing based on device data improves compression quality and treatment outcomes, so recording the event and the ability to read it after the call are essential functions of the device.

Whether the pauses in compressions were shorter than 5 seconds can only be assessed after the intervention if the device recorded the sequence of events. Without a recording, the debriefing relies solely on the team’s memory.

Requirements for a defibrillator in an ambulance

A defibrillator in an ambulance operates in different conditions than in a hospital. It is carried up stairs, exposed to moisture and low temperatures, and must be secured and powered inside the vehicle.

ParameterWhy it is importantWhat to check
IP protection ratingThe device is used outdoors, in rain and dustDeclared IP rating in the manufacturer’s documentation
Drop resistanceThe equipment is carried and may be droppedDrop test height
Operating temperature rangeThe temperature in the vehicle changes with the seasonsLowest and highest operating temperature
Weight with batteryThe paramedic carries the device along with a bagWeight of the set with battery and bag
Power supplyThe battery must last for the interventionNumber of batteries and charging method in the vehicle
MountingUnsecured equipment is a hazard while drivingWhether the bracket is designed for the model and crash-tested

Resistance and weight

An IP55 protection rating means protection against dust and water jets, and a drop test from a specified height shows whether the device will survive a typical accident during the team’s work. Both values are checked in the manufacturer’s documentation.

An example is the parameters of Schiller’s DEFIGARD Touch 7 from the KG Special Performance offer.

ParameterValue
Weight3.3 kg with battery and bag
ModesSemi-automatic and manual
Defibrillation impulseBiphasic, Multipulse Biowave
Maximum energy200 J
PatientsAdults and children
Protection ratingIP55
Operating temperatureFrom 0 to +50 °C
Drop testFrom a height of 1 m

Power supply and charging in the vehicle

An ambulance defibrillator runs on a battery and requires space and a charging method in the vehicle. Before purchasing, you should check whether the ambulance build has a socket and bracket intended for the given model and whether the battery can be charged in the vehicle. In KG Special Performance ambulances, all electrical connectors have an IP44 protection rating and LED indicators.

Mounting and compatibility with other equipment

Mountings in KG Special Performance ambulances are tested for a 10G load. The defibrillator must remain in place during sudden braking or a collision, which is why the bracket is selected for a specific model.

Electromagnetic compatibility is verified for the entire build, not for individual devices. The defibrillator operates alongside other electrical equipment in the medical compartment, so its behavior as part of a complete set is what counts.

The standard for road ambulances is PN-EN 1789+A1:2024-06, implementing EN 1789:2020+A1:2023. Drafts of new parts, prEN 1789-1 and prEN 1789-2, are in development and do not change the current requirements.

How to check a defibrillator for an ambulance before purchase

Before purchasing, you need to compare the device’s functions with the equipment list from the National Health Fund ordinance, the team’s qualifications, and the working conditions in the vehicle. In practice, this comes down to six matters:

  • The regulation on guaranteed emergency medical services does not list individual devices. The minimum equipment list for ambulances can be found in Annex No. 3 to Ordinance No. 179/2020/DSM of the President of the National Health Fund, which has been amended multiple times, so it is checked in the consolidated text.
  • A public contracting authority that describes the subject of the contract by referring to a standard must allow for equivalent solutions and state this in the description. The contractor demonstrates equivalence in the bid, in particular using objective means of proof.
  • The device should support all procedures that the team performs independently, from defibrillation to cardioversion, pacing, and capnography.
  • Conditions in the vehicle include IP protection rating, drop test, temperature range, weight, charging, and mounting.
  • Manual mode requires practiced proficiency, and changing a model in the fleet means retraining the team.
  • Pads and batteries have a limited lifespan, so service and access to consumables must be ensured permanently.

FAQ

Can a paramedic perform cardioversion or pacing without a doctor?

Yes. The regulation allows a paramedic to independently perform cardioversion in tachyarrhythmias and transcutaneous cardiac pacing in bradyarrhythmias in hemodynamically unstable patients. This requires a device equipped with such modules.

Why does an ambulance need ECG teletransmission?

Without it, a paramedic cannot administer clopidogrel, prasugrel, or ticagrelor, because the regulation makes their administration conditional on ECG teletransmission and consultation with the doctor evaluating the reading. For the team, it is a condition for using this authorization; for the doctor, a way to assess the reading remotely.

Can a defibrillator be used while a mechanical chest compression device is operating?

Yes. A shock from a manual defibrillator can be delivered without interrupting mechanical compressions. Defibrillating during manual compressions is not allowed.

Is any portable defibrillator sufficient for an ambulance?

No. Portability does not determine compliance with the National Health Fund list, nor does it guarantee resistance to operating inside a vehicle.

What to look for when mounting a defibrillator in an ambulance?

The bracket must be dedicated to a specific device model and withstand the G-forces during braking and collisions. Accessibility is also key, as the device must be removed quickly, and returned to the vehicle just as quickly after the intervention.

Sources

  1. Resuscitation Council UK, Adult advanced life support Guidelines, 2025.
  2. Minister of Health, Regulation on emergency medical procedures and health services other than emergency medical procedures that may be provided by a paramedic, consolidated text, Journal of Laws 2025, item 576, Annex No. 1.
  3. President of the National Health Fund, Ordinance No. 179/2020/DSM of November 12, 2020, Annex No. 3.
  4. Polish Committee for Standardization, PN-EN 1789+A1:2024-06, implementing EN 1789:2020+A1:2023.
  5. European Committee for Standardization (CEN), prEN 1789-1 rev and prEN 1789-2 rev, draft standards, 2025.
  6. KG Special Performance, DEFIGARD Touch 7, Schiller – product data sheet, 2025.
  7. KG Special Performance, internal data regarding mountings and electrical connectors in ambulances.